MP100H/400H
|
MP100/400H features the HygroClip S3 interchangeable sensor module for high performance, and simplified field maintenance and calibration. The sensor module can be calibrated under laboratory or workshop conditions for best performance, then simply interchanged on site in minutes.
Other Features:
Direct Pt100 temperature measurement for the ultimate in temperature accuracy
Temperature probe can be calibrated in ice-point references, liquid baths etc.
Precise & stable %rh & temperature measurement
Specially designed for outdoor measurements
Cable length compensation up to 100m on voltage output versions
Sensor module calibration via PC and software (MOK-03-WIN, HW3)
|
| Operating range (electronics) |
-40...+60°C |
| Input signal |
DIO digital from HygroClip S3 |
| Output signal |
MP100H-010300 0...1V = 0…100%rh, -40...+60°C |
| Output signal |
MP100H-430300-10DIN: 0...1V = 0...100%rh, °C Pt100 4 wire direct, 1/10 DIN |
| Output signal |
MP400H-480300-10DIN: 4...20mA = 0...100%rh, °C Pt100 4 wire direct, 1/10 DIN |
| Output load |
MP100H: min 1000W; MP400H max. 500W |
| Power supply |
MP100H: 3.6...35VDC, 10mA; MP400H: 5V+(0.2 x load), max 45mA |
| Cable length compensation |
Up to 100m |
| Housing |
IP65 white Delrin probe tube 232 x 25mm Ø |
| Cable |
3m white PVDF cable, stripped ends; other lengths available on request |
| Typical applications include weather stations, airports, ice-warning systems, panorama TV systems, and outside condition monitoring for building management systems. |
| |
|
|