HYGROGEN CHAMBER

 
Description
HG Chamber The HygroGen test chamber or working space is key to the units ability to provide a precise and stable reference that any humidity calibration needs.

The careful design and construction means that we are able to achieve temperature control stability of better than 50 milli Kelvin, and temperature gradients where probes are tested of 0.1°C.

A further unique feature of the HygroGen chamber is the availability of three HygroClip probe connections points. The control probe is installed furthermost into the chamber, and two additional positions in the middle and front mean that internal temperature homegenity can be observed by the user.

The digital output signal from each probe is available on the rear of the HygroGen to be connected to either a HygroPalm or HygroLab display instrument using the connecting cable B5-02-B5.
 
Technical Data
Control Stability > ± 0.05 °C
Temp. Gradients (at 23°C) > ± 0.1 °C
Temp. Gradients (at 10°C) > ± 0.2 °C
Temp. Gradients (at 50°C) > ± 0.15 °C